    TPMJW – аппарат теплопроводности с Ⅲ пластиной (метод измерителя теплового потока)

    Категория:

    Свяжитесь сейчас

    Thermal conductivity (or thermal resistance) is one of the main thermal properties of thermal insulation materials, the measurement of thermal conductivity of building materials can be divided into two categories, steady state method and unsteady state method. The various forms of the two methods have their own characteristics and applicable conditions, different materials according to their own characteristics and use conditions, can choose different methods to determine. The method based on steady-state heat conduction principle has been very mature at home and abroad. At the end of 80\'s, China has developed a series of national standards with reference to international standards. Protective hot plate method and ball method thermal conductivity measurement instrument has also been put into production. However, rapid, simple, high accuracy, moderate price of heat flow meter thermal conductivity meter has not been officially produced, the company is specialized in the development and production of thermal testing instruments, especially the use of microcomputer technology for measuring instruments, has more than ten years of history. According to the international GB10295, the jw-iii building material heat flow meter with single sample and double heat flow meter heat conductivity meter developed adopts intelligent measurement. Therefore, the instrument is easy to use and the error caused by human is reduced.
    The instrument was calibrated using NBS SRM 1450b fiberglass board, and several designs with thermal resistance greater than 0.2m·k/W were selected for comparison with the protective hot-plate device, with a deviation of less than ±3%. Therefore, the instrument is suitable for scientific research, design and production units.

    1. Product introduction
    Thermal conductivity (or thermal resistance) is one of the main thermal properties of thermal insulation materials, the measurement of thermal conductivity of building materials can be divided into two categories, steady state method and unsteady state method. The various forms of the two methods have their own characteristics and applicable conditions, different materials according to their own characteristics and use conditions, can choose different methods to determine. The method based on steady-state heat conduction principle has been very mature at home and abroad. At the end of 80's, China has developed a series of national standards with reference to international standards. Protective hot plate method and ball method thermal conductivity measurement instrument has also been put into production. However, rapid, simple, high accuracy, moderate price of heat flow meter thermal conductivity meter has not been officially produced, the company is specialized in the development and production of thermal testing instruments, especially the use of microcomputer technology for measuring instruments, has more than ten years of history. According to the international GB10295, the jw-iii building material heat flow meter with single sample and double heat flow meter heat conductivity meter developed adopts intelligent measurement. Therefore, the instrument is easy to use and the error caused by human is reduced.

    The instrument was calibrated using NBS SRM 1450b fiberglass board, and several designs with thermal resistance greater than 0.2m·k/W were selected for comparison with the protective hot-plate device, with a deviation of less than ±3%. Therefore, the instrument is suitable for scientific research, design and production units.

    2. standard of execution
    ГБ/Т10295-2008
    3. Main technical parameters
    Measurement range: sample thermal resistance >0.1m·k/W
    Lambda value: 0.001 ~ 0.5
    Hot surface temperature: room temperature ~60℃
    Cold noodle temperature: room temperature ~5℃
    Sample size: 300×300mm
    Power supply voltage: AC220V2kW
    Environmental conditions: temperature 5~40℃, relative humidity <90%
    Measurement error: ≤±5%

    Запросить запрос

    О нас

    Компания Beijing Tihoo Technology Co., Ltd. в основном занимается разработкой, производством и продажей различных видов строительных материалов и оборудования для инженерных испытаний. В последние годы было разработано и постепенно выведено на рынок разнообразное испытательное оборудование с независимыми правами интеллектуальной собственности. Продукция, запущенная в производство, включает энергосбережение зданий, строительные двери, окна и навесные стены, строительные материалы и муниципальное испытательное оборудование.

    Запросить цену

    Связаться с нами

  • Пекинская компания Tihoo Technology Co., Ltd.
  • +86 18501073979tianhao-caiqingqing@cabr-th.com
  • Комплексное здание Цзянькеян, Северное Третье восточное кольцо, район Чаоян, Пекин.
  • Copyright © 2026 Пекинская компания Tihoo Technology Co., Ltd. Все права защищены.политика конфиденциальности

    Связаться с нами

    Запросить запрос